I N S T R U C T I O N
, L O A
D I N G O F D A R T S Y R I N G E
Loading
of injection rifle and pistol dart syringe.
In
the interests of safety these instructions must be strictly followed. It
is strongly suggested that the filling process be first practised with
water and repeated several times until you become familiar with it.
Remove
the hypodermic needle from the protective case. Slide the green/red
silicone sleeve onto the needle so that the injection holes are located at
the centre of the sleeve. This can be achieved by rotating the needle
whilst exerting pressure on the sleeve.
The green/red silicone sleeve
must only be used once.
Remove
the red stabiliser from the dart if necessary using a DAN-INJECT venting pin,
release any retained air from the dart syringe air chamber.
Hold
the dart with the air chamber uppermost. Using an air filler
syringe fitted with a DAN-INJECT coupling adapter connected to the drug
chamber, position the black plunger at the rear of the chamber.
Reverse
the dart so that the drug chamber is now uppermost. Using a suitable
syringe filled with the required tranquillising/medicating liquid slowly
inject the drug into the chamber. Ensure that you use a sufficiently small
gauge needle to allow air from the drug chamber to be expelled without
displacing the drug.
Mount
the hypodermic needle onto the dart syringe boss using pliers and locate
it firmly by rotating it slightly whilst applying pressure.
Apply a safety cap over the needle and seat it firmly on the dart.
Pressurising
the dart: Hold the dart horizontally with the needle/protection
cap
uppermost. Mount the
DAN-INJECT coupling adapter onto a dry standard 12 -
20 ml syringe. Introduce 12 ml of air into the syringe, (correct air pressure
for a 3 ml dart). Connect the syringe securely to the
air chamber of the
dart and, with a smooth continuous action, inject the
12 ml of air into the
dart air chamber.
Now 12 ml of air is let in to the syringe though the
coupling. The red plunger will act as a non-return valve and retain the
air within the dart. WARNING: THE DART IS NOWPRESSURISED AND EXTREME CAUTION MUST BE
EXERCISED AT ALL TIMES. SAFETY CRITERIA MUST BE ADHERED TO.